Chapter 3. CT Image Evaluation
1. Spatial Resolution(MTF) evaluation
1.1. Spatial Resolution (MTF) standards
Parameter LSL (Lower Specification Limit) USL (Upper Specification Limit)
MTF 10% <lp/mm> 1.00 3.50
1.2. Phantom Specification
QUART: DVT Phantom