EasyManuals Logo

Vatech A9 Technical Manual

Vatech A9
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #203 background imageLoading...
Page #203 background image
Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
202 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
031 - This error occurs while moving the
origin of the right side of the 4-axis
collimator.
Motor 4 Collimator right
axis error
032
- This error occurs while moving the
origin of the 4-axis collimator Up.
Motor 4 Collimator up axis
error
033 - This is an error that occurs while
moving the 4-axis collimator down the
origin.
Motor 4 Collimator down
axis error
034
- This error occurs while moving the 2-
axis collimator UP/Down origin.
Motor 2 Collimator
up/down axis error
035 - This error occurs while moving the
origin of the Left/Right 2-axis collimator.
Motor 2 Collimator left/right
axis error
036 - This error occurs while moving the
origin of the Left/Right 1-axis collimator.
Motor 1 Collimator left/right
axis error
037
- This error occurs during Generator
Tilting.
Motor generator tilting axis
error
039 - This error occurs while moving the
origin of the X-axis motor.
Motor X-axis error
040
- This error occurs while moving the
origin of the Y-axis motor.
Motor Y-axis error
041 - This error occurs while the sensor X-
axis motor origin is moving.
Motor Sensor X-axis error
003 ---- [ Exposure Switch related error ]
000 - Unknown error
060 - This error indicates that the exposure
switch is pressed when the device is
turned on.
EXP Push error
061 - This is an error that stops X-Ray
exposure by releasing the exposure
switch.
004 ---- [ Other errors ]
000 - Unknown error
100
- Headrest error. In PANO mode, the
headrest must be removed.
Headband support error
101 - Hand frame error. The hand frame is
open. Equipment stops running.
Hand frame support error
102
- This error occurs when CAN
communication is not responding.
CAN communication error
103 - This error occurs when the column
position sensor is not responding.
Column position sensor
error
104 - This error occurs when the column is
not automatically adjusted.
Column auto control error

Table of Contents