EasyManua.ls Logo

Vatech A9

Vatech A9
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
91 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
7. When 8.5mGy is measured at 95kVp, input the mA value to PwrParam_CBCT.ini.
(C:\VCaptureSW\Acquisition\CBCT\WidePANO\UserParam\PwrParam_CBCT.ini)
8. If 8.5 mGy is measured at 95 kVp 7.3 mA, input all mA values in the above exposure condition
table as +30 increments.
9. Compare with the DAP value output from the Console Software and check if it is within 15%.
10. If the output DAP of the Console SW is more than 15% different from the measured DAP, correct
the DAP Level in the Control Panel so that the difference is within 15%.
11. DAP levels can be modified in three levels: Low, Normal, and High.
12. Since kVp values are the same, choose High, Normal, or Low depending on mA.
MAN_HARD = kVp, mA
MAN_STAN = kVp, mA
MAN_SOFT = kVp, mA
- RANGE_kVp, RANGE_mA, kVp values are fixed
- Input by changing only mA that meets the standard

Table of Contents

Other manuals for Vatech A9

Related product manuals