4. Imaging System Overview
36 Green Smart User Manual
4.6.3 Emergency Stop Switch
During operation, the following emergency situations may occur:
X-ray emission even after the Exposure Switch has been released
Physical injury to the patient or damage to the equipment
Other emergency situations
If a problem occurs during image acquisition, press the red Emergency Stop Switch to
immediately stop the moving parts and cut off all power to the equipment’s electrical
components. To restart the equipment, turn the Emergency Stop Switch clockwise until it
pops up.
The Emergency Stop Switch is located at the bottom of the Vertical Frame.
4.6.4 Exposure Switch
The Exposure Switch allows the operator to control image acquisition from outside of the
X-ray room.
Press and hold the Exposure Switch down until acquisition is completed. Premature
release of the Exposure Switch will abort image acquisition.
Pressing the Exposure Switch activates the LED indicator to turn yellow. This color
indicates that the X-ray is being emitted.
The Exposure Switch is detachable. Ensure the Exposure
Switch cable is not detached from the unit accidentally during the
operation.
Keep vocal/visual contact with the patient during exposure. If any
problem occurs during exposure, release the Exposure Switch
immediately.