13. Appendices
135
I
PaX-i3D Smart
English
13.2
X-ray Dose Data
The X-ray dose data is extracted from the X-ray Dose Test Report for the
PaX-i3D Smart.
X-ray Dose Test Report for the PaX-i3D Smart maintains dosemetric evaluation of
VATECH dental diagnostic system meets all requirements specified in the IEC Collateral
Standard. To limit unnecessary exposure to the patient, operator or other staff, the PaX-
i3D Smart is designed to comply with IEC 60601-1-3 Part 1 General Requirements for
Safety.
Test Condition
Brand Name (Model)
PaX-i3D Smart(PHT-30LFO)
Sensor type
PANO/CT: Xmaru 1404CF
CEPH: Xmaru 2301CF(Scan Type)
X-ray generator
DG-07C11T2
13.2.1
DAP Table
Test Equipment
Instrument Manufacturer Model S/N
Dose Meter Piranha 205 CB2-06100053
Mode Tested: PANO Normal Adult 13.6s
mA
kVp
6 8 10
[mGy]
[mGyㆍ
cm2]
[mGy]
[mGyㆍ
cm2]
[mGy]
[mGyㆍ
cm2]
60
6.25 40.50 9.33 59.71 1260 81.65
70
9.29 60.20 12.37 79.17 15.64 98.75
80
12.57 81.45 15.65 100.16 18.92 122.60
90
15.33 99.34 18.41 117.82 21.68 140.49