13. Appendices
138
I
PaX-i3D Smart
CT Mode
Test Condition
Tested Mode
CT FOV 100 x 85 18s Low Dose
Applied Tube Voltage Peak [kVp]
99
Applied Tube Current [mA]
14
Mode
Direction [°]
Ultra Low Dose, FOV 100 x 80 5.9s [mR/hr]
1 m (3.3 ft) 1.5 m (4.9 ft) 2 m (6.6 ft)
0 588.2 135.3 87.1
45 549.3 249.4 106.8
90 472.6 307.3 78.4
135 458.8 287.6 89.3
180 12.9 4.6 1.3
225 410.5 288.7 98.2
270 663.2 301.4 112.4
315 429.7 194.2 92.3