63138-US-200110 31
Vertical Pipeline, 30° Inclined, diameter 1.97…3.94
in (
50…100 mm
)
For small pipe diameters 1.97 … 3.94 in (50 … 100 mm), a diagonal radiation path
is
recommended.
The distance of the beam through the medium is thus longer
and an improved measuring effect is achieved.
In such cases, the optional lead
shielding for the detector is recommended to avoid influence from any secondary
radiation sources.
Figure 4.4
Vertical Pipeline, 30° inclined, diameter 1.97 … 3.94
in (
50 … 100 mm
)