37
10 Diagnostics and servicing
VEGAPULS 11 • Two-wire 4 … 20 mA
58350-EN-200806
Code
Text message
Cause Rectication
F105
Determine measured
value
The instrument is still in the start phase,
the measured value could not yet be de-
termined
Wait for the end of the switch-on phase
Duration up to 3 minutes depending on
the measurement environment and pa-
rameter settings
F260
Error in the calibration
Checksum error in the calibration values
Error in the EEPROM
Send instrument for repair
F261
Error in the instrument
settings
Error during setup
False signal suppression faulty
Error when carrying out a reset
Repeat setup
Carry out a reset
F265
Measurement function
disturbed
Program sequence of the measuring func-
tion disturbed
Device restarts automatically
Code
Text message
Cause Rectication
C700
Simulation active
A simulation is active Finish simulation
Wait for the automatic end after 60 mins.
Code
Text message
Cause Rectication
S600
Impermissible electron-
ics temperature
Temperature of the electronics in the non-
speciedrange
Check ambient temperature
Insulate electronics
S601
Overlling
Dangerofvesseloverlling Makesurethatthereisnofurtherlling
Check level in the vessel
S603
Impermissible operating
voltage
Terminal voltage too small Check terminal voltage, increase operat-
ing voltage
Code
Text message
Cause Rectication
M500
Error in the delivery
status
The data could not be restored during the
reset to delivery status
Repeat reset
LoadXMLlewithsensordataintothe
sensor
M501
Error in the delivery
status
Hardware error EEPROM Send instrument for repair
M504
Error at a device inter-
face
Hardware defect Check connections
Exchanging the electronics
Send instrument for repair
Function check
Outofspecication
Maintenance