Operations & Maintenance Manual: Product Specications
0900-5165 REV-00
Electromagnetic emissions
Table 13. Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
The system is intended for use in the electromagnetic environment specied below. The customer or the
user of the system should ensure that it is used in such an environment.
EMISSIONS TEST COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT –
GUIDANCE
RF emissions
CISPR 11
Group 1
The system uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The system is suitable for use in home
healthcare and professional hospital
environments.
Harmonic emissions
IEC 61000-3-2
Class B
Voltage uctuations/icker emissions
IEC 61000-3-3
In compliance
Accessory Conformance to Standards
To maintain electromagnetic interference (EMI) within certied limits, the system must be used with the
cables, components, and accessories specied or supplied by Verathon. For additional information, see
System Parts & Accessories. The use of accessories or cables other than those specied or supplied may
result in increased emissions or decreased immunity of the system.
Table 14. EMC Standards for Accessories
ACCESSORY MAX CABLE LENGTH
Monitor power adapter 1.5 m (4.9 ft)
Charging cradle power adapter 1.5 m (4.9 ft)