Page | 12
VORTEX AIR Technical Manual
Revision 6- Issued by N. Shields 24/08/2023
3.4. Electrical Specifications
The machine is supplied with a 3-meter length of 1.5mm flex, which shall be connected to either:
230 Volt Machine 10 Amp MCB to BS EN60898 Type C or alternatively fused 16 Amp to BS
EN60269-1. An appropriate I.E.C Approved 16 Amp isolator shall be installed in both cases.
Current rating 8 Amp. Frequency Range 50/60Hz
110 Volt Machine 16 Amp MCB to BS EN60898 Type C or alternatively fused 16 Amp to BS
EN60269-1. An appropriate I.E.C Approved 16 Amp isolator shall be installed in both cases.
Current rating 9 Amp. Frequency Range 50/60Hz
Pollution degree 2, Overvoltage category 11
It is recommended that the disposal unit is supplied by a dedicated circuit.
Where supplementary protection is provided by an RCD/RCBO, this should be rated at 100mA.
Note that the machine should not be switched off at the mains supply mid-cycle except in the
case of an emergency. Doing so will impair the effective operation of the machine and may
cause damage to the components.
Under NO circumstances should the machine be worked upon and dismantled without fully
isolating the machine from the electrical supply.
This machine complies with current safety requirements. Inappropriate use can, however, lead
to personal injury and damage to property. To avoid the risk of accidents and damage to the
machine, please read these instructions carefully before using it for the first time. They contain
important information on its safety, use and maintenance. Keep these instructions safe place
and ensure that new users are familiar with the content.