High-Precision Tuning Fork
Electronic Balance
AJ-E/AJH-E Series
Operation Manual

SHINKO DENSHI CO., LTD.
IMPORTANT
To ensure safe and proper use of the balance, please read
this manual carefully.
After reading this manual, store it in a safe place near the
balance, so you can review it as needed.

Questions and Answers

  • R
    Robert AtkinsonAug 15, 2025
    What to do if Vibra AJ-E Scales displays [?????] before capacity?
    • L
      Linda AtkinsAug 15, 2025
      If the Vibra Scales display shows '?????' before reaching capacity, it could be due to the gross weight exceeding the scale's capacity. Make sure to check the total weight, considering both the container and the sample. Alternatively, there might be damage to the mechanism.
  • J
    Jocelyn AlexanderAug 18, 2025
    Why does my Vibra AJ-E Scales weight indication contain an error?
    • A
      Ashley CarterAug 18, 2025
      If your Vibra Scales displays an error in weight indication, it could be due to a mistake during the tare subtraction procedure. Review the tare subtraction process. Also, ensure the adjusters are not lifted, as this can cause an incorrect level. Check the level. If the indication values are inconsistent after prolonged use or relocation, execute span adjustment on the balance.
  • H
    hfisherAug 21, 2025
    Why is the display unstable on my Vibra Scales?
    • K
      Kimberly JonesAug 21, 2025
      If the display of your Vibra Scales is unstable and [M] keeps flashing, this could be because the balance is exposed to air currents or vibration, or is placed on an unstable surface. Check the precautions on use. It might also happen if an object is touching the sample, the measuring pan, or the tare. Check the precautions on use.
  • A
    alvarezkellyAug 25, 2025
    What does [?????] displayed on Vibra AJ-E mean?
    • F
      fmorrowAug 25, 2025
      If your Vibra Scales displays '?????', it could be due to a foreign object caught between the measuring pan (or pan base) and the balance. Remove the measurement pan and examine the surface beneath it. It might also indicate damage to a section of the mechanism. Remove the measurement pan and examine the surface beneath it.
  • M
    Maria GregoryAug 27, 2025
    Why does limit function not work on Vibra AJ-E Scales?
    • M
      Megan ParkerAug 27, 2025
      If the Vibra Scales limit function is not working, it could be because the function is not selected. Check the operation of the limit function. Alternatively, the limit value might have been entered incorrectly. Check the operation of the limit function.
  • E
    Eric SchmidtAug 30, 2025
    How to fix no display on Vibra Scales?
    • N
      Nathan BurtonAug 30, 2025
      If your Vibra Scales shows no indication on the display, make sure that the AC adapter is properly connected.
  • R
    Raymond NicholsonSep 2, 2025
    What does [?????] displayed during span adjustment (AJ-E) on Vibra Scales mean?
    • T
      Tyler GonzalesSep 2, 2025
      During span adjustment (AJ-E) if '?????' is displayed on your Vibra Scales, it might be because a weight heavier than the capacity was used. Check that the span adjustment procedure was performed correctly. It could also mean that the reference weight is less than 50% of the capacity. Check that the span adjustment procedure was performed correctly. Another possibility is that the calibration produced an error of 1.0% or more. Check that the span adjustment procedure was performed correctly.
  • C
    Charles Young Jr.Sep 5, 2025
    What to do if [???] appears on Vibra AJ-E Scales?
    • M
      Melissa PatelSep 5, 2025
      If '[???]' appears on your Vibra Scales (with '[?]' and a value flashing at '[LO]'), it's likely due to insufficient sample weight in the counting mode. To resolve this, execute the Memory Update Method.
  • K
    Keith MillerSep 8, 2025
    What does [?????] displayed at span adjustment (AJH-E) on Vibra AJ-E Scales mean?
    • T
      thomas67Sep 8, 2025
      If at span adjustment (AJH-E) '?????' is displayed on your Vibra Scales, it may be because the adjustment was performed with something loaded on the measurement pan. Check the correct span-adjustment procedure.
  • J
    julie67Sep 12, 2025
    Why does Vibra AJ-E indication disappear during battery installation?
    • S
      Scott GomezSep 12, 2025
      If during battery installation the indication disappears, '[ ]' flashes, or no indication is produced on your Vibra Scales, it could be because the automatic power-off function was activated. Switch the power on again and deactivate the automatic power-off function if necessary. Alternatively, the battery capacity might be low; recharge the battery.

Summary

Precautions on the Use

Names of Component Parts

LCD Indicators and Operating Keys

Details the balance's display indicators and their functions.

Names and Functions of Operating Keys

Describes the purpose and operation of each key on the balance.

Basic Operations

Installation

Step-by-step guide for setting up the balance correctly.

Functions

Setup and Checking of Functions

Guide on configuring and verifying balance functions.

Switching Function for Units of Measurement

Counting Pieces

Measuring Percentage

Limit Function

Limit Function Setup

Steps to configure the limit function for measurement discrimination.

Setup of Limit Values by Actual Quantity Loads

Setting limit values by weighing actual samples.

Setting up Limit Values by Inputting Values

Entering limit values numerically using the balance keys.

Calibrating the Balance

Calibration of AJ-E Series

Procedure for calibrating the AJ-E series balances.

Calibration of AJH-E Series

Procedure for calibrating the AJH-E series balances.

Input;output functions

Operating the Balance with the Battery

Troubleshooting

Specifications

Vibra AJ-E Series Specifications

General IconGeneral
Stabilization Time2 sec
InterfaceRS232C
Unitsg, lb, oz
CalibrationExternal Calibration
MaterialStainless Steel (Pan)
Platform SizeØ128mm, 170(W) x 140(D)mm