EasyManua.ls Logo

Vibra AJ-E Series - Operating the Balance with the Battery

Vibra AJ-E Series
47 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
11. Operating the Balance with the Battery
This function can be used only when the balance is battery-operated.
11.1 Specifications
Built-in nickel-cadmium battery
Charging time: Approximately 12 hours
Drive time: Approximately 24 continuous hours
Number of charge/discharge cycles: 300 or more
11.2 Charging Method
While the balance is battery-operated, [ ] stays on. The indicator flashes [ ] (charging
required) when battery capacity decreases. If the balance flashes [
], charge the battery by
following these steps:
(1) Connect the dedicated AC adapter to the balance.
(2) Turn the balance off.
(3) Charging takes approximately 12 hours, with power switched off. Charging the battery longer than
12 hours decreases battery life.
11.3 User Precautions
1. Once charging is complete, use the balance without the AC adapter to avoid over-charging. This
can occur since the balance continues to charge the battery with a weak current when the power is
switched on. Overcharging will also decrease battery life
.
2. When the balance is used for the first time after purchase, the operating time may be shorter than
when using a fully charged battery. This is due to natural discharge of the battery. Although the
balance can be used while [
] is flashing, it should be recharged as soon as possible.
3. When the battery displays no indication, or an indication disappears quickly after the balance is
switched on, battery capacity is low. In these cases, either charge the battery immediately or plug
in the AC adapter.
4. Charging the battery while [
] is displayed reduces battery life.
!
Cautions To operate the balance safely, observe the following (failure to do so could result in
malfunctions, breakage, burst batteries, or fire):
1. Do not disassemble or modify the battery. Do not reverse the balance connection or short-circuit
the positive and negative polarities of the balance.
2. Use only the supplied AC adaptor.
3. Do not incinerate used batteries. Dispose as hazardous material only.

Related product manuals