Instruction for tare subtraction (zero adjustment)
C1 C2 Code Description Value Response
T (SP) 54H 20H
Instruction for tare
subtraction (zero
adjustment)
None
A00: Normal end
E01: Tare subtraction cannot be
executed due to an error in
the weight value.
(3) Setup of output control
C1 C2 Code Description
O
0 4FH 30H Stop output
O
1 4FH 31H Output continuous at all times
O
2 4FH 32H Output continuous if stable (stop output if unstable)
O
3 4FH 33H
Outputs once by pressing Memory key (irrespective of whether
stable).
O
4 4FH 34H Outputs once if stable. Outputs if the balance is stable when a
sample is loaded after the preceding sample has been removed
and the balance indicated zero, or less.
O
5 4FH 35H Outputs once if stable, and stops output when unstable. Even if
the sample is not replaced, the balance is output once when it
stabilizes next time (including the zero indication).
O
6 4FH 36H Outputs once if stable, and outputs continuously when unstable.
Even if the sample is not replaced, output of the balance stops
when it stabilizes after being output once.
O
7 4FH 37H
Pressing Memory key causes the balance to output once when
stable.
O
8 4FH 38H Output once immediately.
O
9 4FH 39H Output once after stabilization.
※The output controls executed with commands [O0] - [O7] work the same as the output controls
executed through function setup on the main unit of the balance.
The commands [O8] and [O9] are data request commands issued to the balance.
※Once any command from [O0] to [O9] is executed, the balance runs that function until another
command is entered. However, if the balance is switched off and on again, the output control is reset
to the initial function (function set value).
10.5.3 Response Output
(1) Response output format
Composed of five characters, including the terminators (CR = 0DH; LF = 0AH)
1 2 3 4 5
A1 A2 A3 CR LF
(2) Types of response outputs
A1 A2 A3 Code Description
A 0 0 41H 30H 30H Normal end
E 0 1 45H 30H 31H
Command error
(Abnormal command received; other errors)