OPERATION
Scale
` Use the
and buttons to set the size of the calibration piece that
can be viewed on the monitor that will be used for calibration.
Right calibration line
` Use the and buttons to set the position of the right calibration line
as displayed on the screen.
Left calibration line
` Use the and buttons to set the position of the left calibration line
as displayed on the screen.
Calibration
` Press
to activate calibrated measurements.
` CAL will be displayed in top left corner of the Status bar
` Features that affect the calibration will be disabled while the calibration is
active. Press
again to turn this feature off to reactivate disabled features.
Calibration
Before setting a calibration, ensure the unit is in Manual Focus
mode at Inf. When a calibration is set, the unit is automatically set
to Manual Focus mode and this can cause the object to move out of
focus if working in Automatic Focus mode.
` With the calibrate options ( ) displayed, use the and buttons
to highlight the option required and then press to enable the value
to be altered if necessary.