Vitrek V4 – Electrical Safety Tester USER MANUAL
12
Ramp time and test time
AC: Ramp/Test (000.0~999.9 s)
DC: Ramp/Test (000.0~999.9 s)
IR: Test (001.0~999.9 s)
GB: Test (000.0~999.9 s)
Ramp/Test
Time
CNT: Test (000.0~999.9 s)
CHANNEL
Hi: 00
Lo: 00
Any of the channels can be selected from the Scanner Box for output. If the selection
are Hi=01 and Lo=02 means channel 1 is high voltage output, and channel 2 is at
Return or Source terminal.
When the measured load becomes capacitive, the test time must be extended.
5-2. Prepare the EST for Use
To view the Storage Steps
1. Press the MENU key to enter status MENU.
2. Use the left and right arrow keys to change knob’s resolution (group or step). Use the knob to change the
active step.
voltage
ramp
test
time
MENU
MENU
0-00 AC