Vitrek V4 – Electrical Safety Tester USER MANUAL
19
5-4. Menu Parameter Setup
z AC/DC Hipot – Voltage withstand test (ACW/DCW)
Press MENU key to enter MENU mode then use knob and arrow keys to select a step.
Press EDIT/SAVE key to enter EDIT mode. The cursor stays at the “test mode” field. Use knob to
select desired test ACW (DCW).
Functionality of FIELD key:
Press FIELD key to edit the next field “output voltage”. Use the knob to adjust the desired output
voltage while using the arrow keys to adjust the knob’s resolution (0.100~ 5.000 kV for ACW, 0.100~
6.000 kV for DCW).
Press FIELD key again to enter the next field “measurement upper limit”. Use the knob to adjust the
desired upper limit of leakage current while using the arrow keys to adjust the knob’s resolution
(0.10~15mA for ACW, 0.10~7.5mA for DCW).
Press FIELD key again to enter the next field “measurement lower limit”. Use the knob to adjust the
desired lower limit of leakage current while using the arrow keys to adjust the knob’s resolution
(0.10~15mA for ACW, 0.10~7.5mA for DCW).
Press FIELD key again to enter the next field “ramping time”. Use the knob to adjust the desired
ramping time while using the arrow keys to adjust the knob’s resolution (0~999.9s).
Press FIELD key again to enter the next field “testing time”. Use the knob to adjust the desired testing
time while using the arrow keys to adjust the knob’s resolution (0~999.9s).
Press FIELD key again to return to the first field “test mode” again.
Press EDIT/SAVE key to save all the parameters.
z Continuous AC/DC withstanding voltage test (CAC/CDC)
The CAC/CDC test is available only on “COM”. Like the traditional hipot tester, you can use the knob
and arrow keys to adjust output voltage during the test.
All the parameters of CAC/CDC are the same as ACW/DCW, except the testing time. The testing of
CAC/CDC is not limited.
Test mode
ACW (DCW)
Output voltage
V=X.XXXkV
Measurement upper limit
Imax=XX.XxmA
Measurement lower limit
Imin=XX.XxmA
Ramping time
RAMP=X.XXXs
Testing time
TEST=X.XXXs