Vitrek V4 – Electrical Safety Tester USER MANUAL
35
Enable Register (SRER).
*STB? Reads Status Byte Register (SBR).
:STATus:OPERation:CONDition ? Returns the contents of the OPERation
condition register. Returns NR1.
:STATus:OPERation:ENABle <NR1> Sets the contents of the enable mask for the
OPERation event register.
:STATus:OPERation:ENABle ? Returns the contents of the enable mask for
the OPERation event register. Returns
NR1.
:STATus:OPERation:EVENt ? Query the contents of the OPERation Event
register.
:STATus:PRESet Presets the OPERation and QUEStionable
status registers.
:STATus:QUEStionable:CONDition ? Returns the contents of the OPERation
condition register. Returns NR1.
:STATus:QUEStionable:ENABle
<NR1>
Sets the contents of the enable mask for the
QUEStionable enable register.
:STATus:QUEStionable:ENABle ? Query the contents of the Questionable
Enable register.
:STATus:QUEStionable:EVENt ? Query the contents of the QUEStionable
Event register.
Table 4: Miscellaneous Commands
Command Explanation
*IDN? Returns instrument identification.
*OPC Reports when operation is complete by setting the Operation
Complete bit in SESR.
*OPC? Reports when operation is complete. Same as *OPC except
returns a 1 to the output queue and dose not set the SESR bit.
*RCL Recall the setting data from the memory which previous saved.
*RST Resets the protection levels and states, resets the current and
voltage levels to zero, sets the output off, and sets memory
section to 00.
*SAV Saves the setting data to memory.
*TST? Initiates internal self-test and reports results.
*WAI Wait to continue. This command forces sequential operation of
commands. This command is required by IEEE-488.1-1987.
The power supply, however, forces sequential operation of
commands by design.