Vitrek V4 – Electrical Safety Tester USER MANUAL
51
:GBONd:CURRent
Function:
Set or query the GB test current.
Syntax:
:GBONd:CURRent <NR1>
:GBONd:CURRent?
<NR1> Please refer to the specification
Return:
<NR1>
Examples:
:GBONd:CURRent - 25 sets the GB test current to 25A.
:GBONd:CURRent? - returns “25” if the GB test current setting is 25A.
:GBONd:RMINimum
Function:
Set or query the GB test minimum resistance.
Syntax:
:GBONd:RMINimum <NR1>
:GBONd:RMINimum?
<NR1> Please refer to the specification
Return:
<NR1>
Examples:
:GBONd:RMINimum100.0 - sets the GB test minimum resistance to 100m.
:GBONd:RMINimum? - returns “100” if the GB test minimum resistance setting is 100m.
:GBONd:RMAXimum
Function:
Set or query the GB test maximum resistance.
Syntax:
:GBONd:RMAXimum <NR1>
:GBONd:RMAXimum?
<NR1> Please refer to the specification
Return:
<NR1>
Examples:
:GBONd:RMAXimum 100.0 - sets the GB test maximum resistance to 100.0m.
:GBONd:RMAXimum? - returns “100” if the GB test maximum resistance setting is 100m.