Vitrek V4 – Electrical Safety Tester USER MANUAL
54
Figure 8: Status registers and related commands
The CONDition register is a read-only register which monitors the present state of the instrument. The
CONDition register updates in real time and the inputs are not latched or buffered. When a condition
monitored by the CONDition register becomes true, the bit for that condition also becomes true (1). When the
condition is false, the bit is 0.
The read-only EVENt register latches any false-to-true change in condition. Once the bit in the EVENt
register is set, it is no longer affected by changes in the corresponding bit of the CONDition register. The bit
remains set until the controller reads it. The command *CLS (Clear Status) clears the EVENt register.
QUEStionable Status Registers.
Table 4: 16 bit QUEStionable Status Register
Bit 15 Bit 14 Bit 13 Bit 12 Bit 11 Bit 10 Bit 9 Bit 8
NU
NU NU NU NU NU NU
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
NU NU NU NU NU NU NU NU
The command STATus:QUEStionable:CONDtion? Reads the QUEStionable CONDition register but dose
not clear it.
The command STATus:QUEStionable:EVENt? Reads the QUEStionable EVENt Status register and clears it.
OPERation Status Registers
Table 5: 16 bit OPERation Status Register
Bit 15 Bit 14 Bit 13 Bit 12 Bit 11 Bit 10 Bit 9 Bit 8
NU NU NU NU NU NU NU
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
NU NU NU NU NU NU NU NU
Status Registers
NU: not used
Enable
Register
Event
Register
Condition
Register
To SBR