Vitrek V4 – Electrical Safety Tester USER MANUAL
6
Ground Bond Test Time 000.0~999.9s
6) ARC Detect
Detect Current 40 level (1~40mA)
7) Storage
Groups
10
Steps
16
8) Interface
RS-232 Standard
GPIB Option
9) PLC Control
D-sub 9 pins female Standard
10) Scanner Interface
D-sub 9 pins female Standard
11) Real Plate Output
Hi-pot terminals and GB
terminals
Standard interface for V4
12) General
Power Source
AC100V, 120V, 220V, 230V±10% 50/60Hz
Operation
Environment
Indoor use, altitude up to 2000m.
Ambient Temperature 0 to 40.
Relative Humidity 80% (Maximum).
Installation category II
Pollution Degree 2
Storage temperature
& Humidity
-10 to 70.
70% (Maximum).
Accessories
TL-20-50 × 1, TL-463 × 1,
Instruction manual × 1,
Power Cord x 1
Dimension 446(L) × 330(W) × 149(H) (m/m)
Weigh
Approx. 14.9 kgs, 35lbs
The main purpose provided by the series of instruments is for Puncture Testing. The specification is not for
continuous test. The temperature of heat sink is monitored. The test procedure will stop if the heat sink is too
hot. Please refer to Figure 2-1.
Stop the instrument for 10 minutes when continuously proceeding 30~40mA withstanding test for one hour.
GENERAL SPECIFICATIONS:
Test Database: 10 Test Sequences with up to 16 steps per test sequence
Front Panel Lockout: Prevents database edit while in lockout mode
Arc Detect: 10 level trip select
Remote Interface: Fully interactive RS232 standard, GPIB optional, standard 9-pin D-sub connector with
Outputs of Testing, Pass, Fail and inputs of Start, Reset. Output isolation is 400V with 130mA Continuous
load current.
Timer Accuracy: 0.10% ± 50mS
Power Source: AC100V, 120V, 220V, 230V (±10% 50/60Hz), 250W max