EasyManua.ls Logo

VOLTCRAFT VC1008 - Resistance measuring;diode test;continuity test; Resistance and Continuity Measurement; Resistance Measurement Precautions; Function Buttons for Resistance;Diode;Continuity

VOLTCRAFT VC1008
104 pages
Go to English
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40
Proceed as follows to measure resistance and for the acoustic continuity test:
- Switch on the power unit on the operating switch (3).
- Select DMM mode with the MODE button.
- Connect the black measuring line with the COM socket
and the red measuring line with the VΩHz socket till
they are plane at the measuring instrument.
- Set the rotary selection switch (3) to Ω.
- Check the measuring leads for continuity by connec-
ting both measuring prods to one another. After that
the resistance value must be approximately 0 Ohm.
- Now connect the measuring prods to the object to be
measured. As long as the object to be measured is not
high-resistive or interrupted, the measured value will be indicated on the display (1).
- As soon as „O.L“ (overflow) appears on the display, you have exceeded the mea-
suring range or the measuring circuit has been interrupted.
- Automatic range selection is activated as default value.
If you carry out a resistance measurement, make sure that the measu-
ring points which you contact with the test prods are free from dirt, oil,
solderable lacquer or the like. An incorrect measurement may result
under such circumstances.
Function buttons F1 – F4
F1 ( ): Switching from resistance measuring ( ) to diode test ( )
or acoustic continuity check ( )
F2 (REL): Relative measuring function – The measuring value is set to 0.000
after pressing the button; now only the relative value to the pre-
vious measuring value is displayed. Press F2 again to deactivate
the relative function.
F3 (RANGE ): Manual range selection - increasing the measuring range. The set
measuring range is indicated on the display. Press AUTO to reac-
tivate the automatic range selection.
F4 (RANGE ): Manual range selection - decreasing the measuring range.
To activate the acoustic continuity tester function, press F1 until ( ) appears. A
measuring value of less than 100 Ohm is identified as continuity; in this case a beep
sounds. Make sure that the beep function is activated (see basic setting).
10 A mA µA
SCOPE
COM
V Hz%

Table of Contents

Other manuals for VOLTCRAFT VC1008

Related product manuals