127V2.0/02.11
Descriptionofthememoryfunctions
ENGLISH
13. Descriptionofthememoryfunctions
Conguringthemeasuredvaluememory
DATA FORMAT ACT
During the recording interval
(for example, 50 ms), the current meas-
urement value (ACT) only will be written
to the measured value memory.
MIN-MAX
During the recording interval
(for example, 50 ms) one MIN and one
MAX value will be written to the meas-
ured value memory.
REC CONFIG 2 000 PTS
The selected recording time is auto-
matically divided into a xed number of
recording intervals per channel.
Example:
10 min recording time = 600 s
Duration of recording interval =
600 s ÷ 2.000 = 300 ms
REC RATE
Denition of an individual recording
interval (for example, 5 ms).
Based on the settings (DATA FORMAT/
REC RATE), the measuring instrument
examines if the selected recording time
must be extended.
Example:
Recording time 100 h/conict
recording time
FAST MODE
ACT measured values only are saved
at a xed recording interval of 0.5 ms
via IN1. All other inputs (INx) are not in
function.