EasyManua.ls Logo

Welch Allyn 420 Series - Pulse Rate Accuracy; Overpressure Cutoff; Temperature Ranges; Oxygen Saturation Range (Spo 2 %)

Welch Allyn 420 Series
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Spot Vital Signs 17
Pulse Rate Accuracy
SpO
2
Module Heart Rate: ±3.0 bpm
Blood Pressure Algorithm Heart Rate: ±5.0%
Overpressure Cutoff
315 mmHg ±15 mmHg
Temperature Ranges
Temperature Accuracy
Calibration accuracy: + 0.2 F (+ 0.1 C).
Temperature Determination Time
Oral: 4 seconds typical, 15 seconds maximum
Axillary: 10 seconds typical
Rectal: 15 seconds typical
Oxygen Saturation Range (SpO
2
%)
40-100% oxygen saturation
SpO
2
Accuracy
±3% in the range of 70-100% oxygen saturation (1 standard deviation)
<70% unspecified by the OEM
Battery Charging
The battery charges to 90%-100% capacity in 12 hours. Unit operates and charges battery simul-
taneously when connected to power source.
Mechanical Specifications
Dimensions
Height: 9.70 inches (24.64 cm)
Length: 5.72 inches (14.53 cm)
Depth: 4.73 inches (12.01 cm)
Weight
Approximately 4.25 pounds (1.91 kg)
Mounting
Self-supporting on rubber feet
Measurement Maximum Minimum
Temperature 109.4° F
43.0° C
86.0° F
30.0° C

Table of Contents