46 Welch Allyn
Maintenance
Welch Allyn will make available,upon request, circuit diagrams and other information which will
assist appropriately qualified technical personnel in repair of this device. Please reference “4200-
145E Service Manual” on page 50.
Cleaning
Spot Vital Signs
Wipe the Spot Vital Signs clean with a cloth slightly dampened with warm water and a mild
detergent solution. Never immerse the Spot Vital Signs in any type of fluid.
Occasionally clean the unit, as necessary, with appropriately diluted, non-staining disinfectant
solution. Use either 70% isopropyl alcohol, 10% chlorine bleach solution, or mild detergent in
water.
Note: Prevent water or other fluids from entering any connectors. Should this occur, dry the connectors
with warm air. Check all measurement functions for proper operation.
Every 3 months, inspect the temperature probe, SpO
2
Cord, and accessories for fraying or other
damage. Replace as necessary.
Do not sterilize or autoclave the Spot Vital Signs.
Blood Pressure Cuff
Clean the durable one-piece blood pressure cuff with a damp cloth, or wash in water with soap or
detergent. Before washing the cuff, remove the tube fitting(s), close off tubes with plugs
(available as accessory 5082-163), and place the hook and loop fasteners in the closed position.
After washing, allow the cuff to air dry. Re-assemble the tube fitting(s).
Disinfection
: You may use glutaraldehyde-type liquid disinfectants on the durable cuff. Prolonged
use of these disinfectants at full strength may cause discoloration of the white cuff markings.
Sterilization:
Do not use steam or heat to sterilize the cuff or tubing. If necessary, use gas ster-
ilization.
Do not press with a hot iron.
Cables and Pressure Hose
Wipe the cabling and pressure hose with a damp cloth moistened in a mild detergent solution.
Do not immerse hose.
Temperature Probe
Periodically clean the temperature probe by wiping with an alcohol-dampened cloth, or wipe with
warm water or properly diluted, non-staining disinfectant. Do not immerse the probe.
SpO
2
Sensor
Clean the reusable SpO
2
sensor with isopropyl alcohol solution, and sterilize it using ethylene
oxide (EtO), cold cycle. Do not immerse the sensor.