6.2.5 Fire hazard
Never use the examination and treatment table Easy Care together with inflammable anaesthesia gas or
other inflammable material. In particular this applies to fluids such as e.g. alcohol, ether etc, which might
ignite.
During operation never cover the radiant warmer with tissues or similar as this would impair the convection.
There is a danger of overheating and spontaneous combustion.
Take care that during operation there is no inflammable material (e.g. curtains) within 50 cm around the
device.
After switch-off the radiant warmer will stay hot for a while. The residual heat is displayed until the radiant
warmer has cooled-off completely to avoid fire and accident.
6.3 Start-up
6.3.1 General
Qualified persons only should install and start-up the examination and treatment table Easy Care and instruct
the user's personnel.
Before start-up the examination and treatment table Easy Care and the connected accessories must be
checked for intactness, correct and secure position as well as for correct function.
Never use a defective device.
Before operation the examination and treatment table Easy Care must be prepared in cold condition
according to chapter 10 "Cleaning and disinfection".
6.3.2 Electrical connection
The data on the name plate must be identical to the
data of the available power supply.
The device must only be connected to a power
supply with earth conductor.
The radiant warmer must be connected with the
cable (C). Turn off the switch (E) on top of the radiant
warmer.