When, according to the room class and application the
examination and treatment table shall be connected to
a protective equipotential bonding, this cable must be
connected to plug (B).
The power cord is passed through the strain relief (A')
and connected to the power plug of the examination
and treatment table Easy Care.
During installation take care that the device plug is
well accessible at all times.
Additional electrical devices can be connected to the
take-off sockets by observing the total power input.
(Refer to chapter 13.4 "Operating / performance data".
When connecting additional devices to the
examination and treatment table Easy Care the
"General requirements for the safety of medical
electrical systems EN 60601-1-1 + Supplement A1" in
their latest version must be observed, in particular the
permissible limit values for the total sum of leakage
currents.
For height-adjustable devices (WY3020) it must be
observed that the whole body of the device, including
the accessories attached to the device, is moved
vertically by an electric motor. Before operation make
sure that no furniture or devices are placed within the
moving range of the height-adjustment**.
For checking the function adjust the patient deck of the
examination and treatment table Easy care with the
foot pedals to the highest and then to the lowest
position.