12. Maintenance
Nomenclature
= Skilled worker, engineer, bio-medical engineer, with corresponding qualification.
Authorized qualified person
= Qualified person, who has acquainted special knowledge of a certain product.
= Ascertainment of the actual state in terms of operational and functional safety of the medical device..
= Measures to maintain the nominal state.
= Measures to restore the nominal state.
= Inspection, preventive maintenance, repair.
Inspection
In order to ascertain the safe operation of the examination and treatment table Easy Care we recommend
inspections by qualified persons at regular intervals which include the following checks.
Verification that the actual use is in conformance with the intended use.
General condition of the device.
Function and secure locking of adjustment facilities (double castors, height-adjustment**)
Functions of the device and of the safety provisions.
Measurement of the electrical safety according to the national standards and limit values actually in force.
Preventive Maintenance
In order to maintain the nominal state of the examination and treatment table Easy Care we recommend an
annual preventive maintenance by authorized qualified persons. In addition to the inspection the preventive
maintenance includes the following checks:
Check of: interior wiring, connections, electrical components, heating element.
Check of all functions and safety-relevant parameters according toEN 60601-1, in particular EN 60601-2-
21.
Weyer GmbH is only responsible for the safety features of the examination and treatment table Easy Care
when preventive maintenance and repairs are carried out by authorized qualified persons, by observing our
instructions and by using original spare parts.
The microprocessor control system recognizes faults and cuts-off the radiant warmer in case of unsafe
conditions. The safety and alarm functions are listed in Chapter 8.
Devices respectively device components must be cleaned and disinfected prior to each maintenance
measure as well as when they are sent to our factory for repair.
12.1 Spare and wear parts
Pad 460 x 720 x 30 mm with upholstery for side walls, for Easy Care
Safety guard for Ceramotherm 3100
LED-band- 5,5 W-175 mA - 850 lm-4100K for Ceramotherm 3100
Lamp cover for Ceramotherm 3100