3. Safety instructions and warnings
The examination and treatment table must only be put into operation when it was duly mounted by
qualified persons and if neither the device nor the fixing elements show any damages. There is a risk of
injury to patient and user.
It is necessary to carefully read the complete instruction for use prior to using the examination and
treatment table. The device must only be operated when the contents of this instruction was understood.
There is a risk of misuse.
The instruction for use must be kept available at all times for users to consult.
The examination and treatment table Easy Care is only suitable for the "intended use" described before
and must only be used under medical supervision by qualified persons. Remarks in this instruction do not
substitute the medical advice. Any other use can cause hazard to patient and user.
To avoid the risk of an electrical stroke this device must only be connected to a power supply with earth
conductor. Furthermore, to avoid electric potential difference, this device should be connected to a
protective equipotential bonding.
In order to completely separate the examination and treatment table Easy Care from the power supply,
the device plug must be disconnected. During installation respectively mounting take care that the device
plug is accessible at all times.
When connecting additional devices to the examination and treatment table Easy Care the "General
requirements for the safety of medical electrical systems EN 60601-1-1 + Supplement A1" in their latest
version must be observed, in particular the permissible limit values for the total sum of leakage currents.
The max. power input for additional devices must be observed (refer to chapter 13 "Technical Data").
In order to ascertain the safe operation and to preserve the value of the examination and treatment table
Easy Care, regular maintenance measures according to chapter 12 are recommended. Weyer GmbH is
only responsible for the safety features of the device when these maintenance measures are carried out
in regular intervals by us or by authorized qualified persons and by using original spare parts.
This device must not be modified without our explicit consent, otherwise any warranty claims and product
liability will be lost.
If the examination and treatment table Easy Care shall be transported or stored, the environmental
conditions from chapter 13.1 must be observed. Shock and strong vibrations can damage the device.