EasyManua.ls Logo

Weyer Easy Care - Operating; Performance Data; Classifikation; Standards; Certification

Weyer Easy Care
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instruction for use Examination and treatment table Easy Care
** Depending on model and configuration
D2151_01 - 09/2017 37/38
13.4 Operating / Performance data
Allgemein
Operating voltage / power supply
230 V~ / 50/60 Hz
Power consumption
WY3019
WY3020
690 W / 3 A
810 W / 3.5 A
Max power input
2300 W / 10 A
Radiant warmer
Distance to the patient deck
800 mm
Heating element
Ceramic, 600 Watt
Wave length spectrum
1500 to 6800 nm
Intensity selection
2 to 20 mW/cm²
Steps of intensity selection
2 mW/cm² respectively 10%
Resolution of the displays
2 mW/cm² respectively 10%
Intensity selection display
LED-spot 2 to 30 mW/cm² respectively 10 to 100%
Actual intensity display
LED bar 2 to 30 mW/cm² respectively 10 to 100%
Residual heat display
Cooling off to 2 mW/cm² respectively 10%,
then automatic switch-off
Safety function (alarm / automatic intensity
reduction)
After 15 minutes above 10 mW/cm²
Heating-up period from 24°C ambient
temperature to max. intensity selection
Approx. 15 minutes
Illumination
2 LED stripes 5 W each
Illumination intensity
2 x 850 lm
Adjustable
5 steps
Colour temperature
4000 Kelvin
13.5 Classifikation
WY3019
WY3020
Protection class
I
I
Type applied part patient deck
B
B
Degree of protection against water
IPX0
IPX0
Classification according to MDD 93/42 EEC,
(Appendix IX, Rule 9)
II a
II a
13.6 Standards
Amongst others the device complies with:
EN 60601-1:2006 +AC:2010 + A1:2013
EN 60601-1-2:2015
EN 60601-2-21:2009 + A11:2011
13.7 Certification
Certification:
MDD 93/42 EEC, Appendix II, excluding Section 4.
EN ISO 13485:2012/AC:2012
Certified body:
TÜV- Rheinland LGA Products GmbH

Table of Contents