Instruction for use Examination and treatment table Easy Care
** Depending on model and configuration
D2151_01 - 09/2017 32/38
Never use inflammable disinfectants !
The examination and treatment table Easy Care is not suitable for cleaning and disinfection
machines or plants.
When using disinfectant concentrates please observe the correct mixing ratio.
Excessively high concentration can lead to material stress or damage.
The exposure time according to the disinfectant manufacturer's instructions must be observed.
10.2 Dismantling, cleaning and disinfection
After use switch-off the radiant warmer and wait until the residual heat display has extinguished. Disconnect
the power plug.
- Remove nappies and tissues.
- Disconnect suction assemblies and patient circuits (if any).
- Empty liquid containers (if any).
Pad:
- Remove the pad from the bed.
- Clean and wipe disinfect pad surface.
Patient deck:
- Remove visible contamination with a soft cloth or disposable cloth soaked in cleansing agent,
- Wipe disinfect the surface and allow the disinfectant to take effect in accordance with the disinfectant
manufacturer’s instructions. Then wipe the surfaces with a soft, slightly damp cloth and finally wipe it
dry.
Castors:
- The castors must be cleaned very carefully and sprayed with a disinfectant.
Radiant warmer:
- Remove visible contamination with a disposable cloth soaked cleaning agent but not drip wet.
- Then disinfect the surfaces by wiping. Take care that the cloth is soaked in disinfectant but not drip wet.
After the exposure time indicated by the manufacturer of the disinfectant wipe the surface with a damp
cloth and then wipe them dry.
No liquid must enter into the radiant warmer
Do not switch on the radiant warmer to accelerate the drying process.
Never threat the ceramic heating elements with cleansing agents or disinfectants.
The examination and treatment table Easy Care is not suitable for cleaning and disinfection
machines or plants.