8. Safety and alarm functions
8.1 Audible alarms and signals
Power failure alarm
High priority alarm
Medium priority alarm
Low priority alarm
Confirmation of selection
Deactivation of automatic intensity
reduction
Reactivation of automatic
intensity reduction
Signal tone Apgar timer
Signal tone countdown timer
8.2 Messages - Errors - Causes - Measures
High priority alams:
Message
Signal tone
Cause / Effect
Measure
ERROR 001
High priority alarm
DMA error
No new data were received
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
Medium priority alarms:
ERROR 006
Medium priority
alarm
EEPROM error
Comparison of the check
sum of EEPROM data failed
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
ERROR 007
Medium priority
alarm
Watchdog error
The test of the external
Watchdog module failed.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked
ERROR 008
Medium priority
alarm
IIC-Bus error
The communication on the
IIC-Bus is disturbed.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.