8. Safety and alarm functions
8.1 Audible alarms and signals
Confirmation of selection
Deactivation of automatic intensity
reduction
Reactivation of automatic
intensity reduction
Signal tone countdown timer
8.2 Messages - Errors - Causes - Measures
DMA error
No new data were received
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
EEPROM error
Comparison of the check
sum of EEPROM data failed
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
Watchdog error
The test of the external
Watchdog module failed.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked
IIC-Bus error
The communication on the
IIC-Bus is disturbed.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.