Instruction for use Examination and treatment table Easy Care
** Depending on model and configuration
D2151_01 - 09/2017 27/38
Relay & Triac error
The safety relay has not cut-
off securely in switched-off
condition or in the case of
fault and heating is triggered
by a Triac or a control defect.
Switch-off the heating and
disconnect the device from the
power supply immediately.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
Minimum performance
error
(Deviation between actual
intensity and selection).
The current flow through the
heating elements is not
sufficient.
Cause:
- Power voltage too low.
- Heating element defective.
- False configuration of
device.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
Maximum performance
error
(Deviation between actual
intensity and selection).
Excess current flow through
the heating elements.
Cause:
- Power voltage too high.
- Heating element defective.
- False configuration of
device.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
Relay error (ON)
The safety relay does not
switch on, although heating
was switched on.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
Relay error (OFF)
The safety relay does not
switch off, although heating
was switched off.
Switch-off the heating and
disconnect the device from the
power supply.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
No capacity
No current flow through the
heating elements
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
Current transducer error
The current transducer was
recognized to be defective.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
Triac error
Caused by a defect the
heating is permanently
triggered.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.