Instruction for use Examination and treatment table Easy Care
** Depending on model and configuration
D2151_01 - 09/2017 27/38
Message
Signal tone
Cause / Effect
Measure
ERROR 009
Medium priority
alarm
Relay & Triac error
The safety relay has not cut-
off securely in switched-off
condition or in the case of
fault and heating is triggered
by a Triac or a control defect.
Switch-off the heating and
disconnect the device from the
power supply immediately.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
ERROR 010
Medium priority
alarm
Minimum performance
error
(Deviation between actual
intensity and selection).
The current flow through the
heating elements is not
sufficient.
Cause:
- Power voltage too low.
- Heating element defective.
- False configuration of
device.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
ERROR 011
Medium priority
alarm
Maximum performance
error
(Deviation between actual
intensity and selection).
Excess current flow through
the heating elements.
Cause:
- Power voltage too high.
- Heating element defective.
- False configuration of
device.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
ERROR 012
Medium priority
alarm
Relay error (ON)
The safety relay does not
switch on, although heating
was switched on.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
ERROR 013
Medium priority
alarm
Relay error (OFF)
The safety relay does not
switch off, although heating
was switched off.
Switch-off the heating and
disconnect the device from the
power supply.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
ERROR 014
Medium priority
alarm
No capacity
No current flow through the
heating elements
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
ERROR 015
Medium priority
alarm
Current transducer error
The current transducer was
recognized to be defective.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
ERROR 016
Medium priority
alarm
Triac error
Caused by a defect the
heating is permanently
triggered.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.