Instruction for use Examination and treatment table Easy Care
** Depending on model and configuration
D2151_01 - 09/2017 28/38
Meldung
Signalton
Ursache / Wirkung
Maßnahme
ERROR 017
Medium priority
alarm
Time limit error
The calculated heating up
respectively cooling off
period has exceeded the
limit.
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
ERROR 018
Medium priority
alarm
Key pad error
At least one key is being
pressed for more than
30 seconds.
Interrupt pressing of the key.
As soon as the error was removed
and reset, the message will
disappear.
When the error persists, switch-off
the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
ERROR 019
Medium priority
alarm
Rotary push-button error
The rotary push-button is
being pressed for more than
30 seconds
Interrupt pressing of the rotary
push-button.
As soon as the error was removed
and reset, the message will
disappear. When the error
persists, switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
ERROR 020
Medium priority
alarm
Real time error
The internal real time clock is
defective or the
communication is disturbed
Switch-off the heating.
Make sure that the patient is
cared for in another way.
Put the device out of operation
and have it checked.
Low priority alarms:
Low priority alarm
Alarm buffer defective
The alarm buffer is defective.
Have the alarm buffer for the
power failure alarm replaced by a
qualified person.
The message will disappear as
soon as the error was removed.
ERROR 027
No alarm
Zero point error
Because of too low power
supply or a device defect the
zero point of the sinusoidal
wave is not recognized.
Possibly the zero point is not
triggered which can cause
errors in the power supply.
The message will disappear
automatically as soon as the
cause was removed.
When the error persists, the
power supply must be checked.
If necessary, put the device out of
operation and have it checked.
ERROR 028
No alarm
Ambient light sensor error
The ambient light sensor is
defective.
Display changes to full
brightness.
The device must be checked.
The message will disappear
automatically as soon as the
cause was removed.