13. Technical data
13.1 Transport and storage
For transport and storage the following conditions should be observed:
Temperature: 0 – 70°C
Relative humidity: 15 to 90 % (no condensation)
Atmospheric pressure 210 to 1100 hPa
Before moving the examination and treatment table Easy Care, it is recommended to adjust it to the lowest
position. Otherwise it may topple over when moving over thresholds and obstacles. Strong vibrations should
be avoided, e.g. moving on cobble stone pavement or similar.
13.2 Operating environment
15 to 90 % (no condensation)
Please also observe chapter 4 "Requirements of the location“
The examination and treatment table Easy Care is not suitable for the use in explosion-hazardous
areas
To avoid the risk of an electrical stroke this device must only be connected to a power supply with
earth conductor.
The examination and treatment table Easy Care must not be operated close to strong or high-
frequent electro-magnetic radiation sources such as HF surgery devices, MRT or similar. Under
unfavourable conditions this could cause faults.