13.8 Remarks to electromagnetic compatibility (EMC)
The examination and treatment table Easy Care complies with the current requirement for the
electromagnetic compatibility EN 60601-1-2:2015. The device is intended for the use in medical areas with
the exception of areas subject to high intensity electromagnetic disturbance, as it could be the case close to
high frequency surgery devices or magnetic resonance tomography (MRT).
When using the examination and treatment table Easy Care in domestic environment, the device might not
provide the necessary protection of communication services. If necessary, the user must take corrective
measures such as relocation or repositioning of the device.
Operating the device close to other devices should be avoided as this could lead to a faulty operation. If such
operation is nevertheless required, then this device as well as the other devices should be checked for
correct operation.
Portable high frequency communication devices as well as their accessories, e.g. antenna cable and
external antennas should not be used beyond a distance of 30 cm to the device and its cables.
Using other accessory components and cables than supplied by the manufacturer of the device, this can
lead to increased electromagnetic interference or a reduced electromagnetic immunity of the device,
resulting in a faulty operation.
If in spite of considering the before mentioned remarks the device is influenced by electromagnetic
interference, the operating parameters of the devices may be impaired. In that case malfunctions of the
heating can occur. Therefore the patient's temperature should be checked in frequent intervals.