Pressure Transducer Calibration (Zeroing the Sensor)
Pressure transducers should be calibrated prior to connection to a machine when taking data.
To calibrate a pressure transducer simply plug the sensor into the analyzer and press the Utilities function key,
then select option 3 (Calibrate sensors) from the menu.
Alternatively, zero points may also be added directly to the data collection route to allow zeroing of the sensor
from the main data collection route menu. By using his option reduces button clicks and prevents the user from
possibly forgetting to zero a sensor if more than one sensor is being used on a machine.
Note: We recommend zeroing the pressure transducer periodically during data collection to compensate for
thermal shift.
Press analyzer function key to open “Utilities”
menu on the analyzer.