Conductivity Cond 7310
22 ba75933e04 06/2013
5 Conductivity
5.1 Measuring
5.1.1 Measuring the conductivity
NOTE
When connecting a grounded PC, measurements cannot be performed in
grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
1. Connect the conductivity measuring cell to the measuring instrument.
The conductivity measuring screen is displayed.
2. If necessary, select the measured parameter with <M>.
3. Check the Measuring cell settings and cell constant for the connected
conductivity measuring cell.
If necessary, correct the settings.
The selection of the measuring cell and the setting of the cell con-
stant is done in the measurement settings menu for conductivity
(see section 6.1.1 C
HANGING THE SETTINGS FOR CONDUCTIVITY MEA-
SURING CELLS, page 28). The cell constant to be set must either be
taken from the operating manual of the measuring cell or is printed
on the measuring cell.
4. Immerse the conductivity measuring cell in the test sample.
5. If necessary, select the measured parameter with <M>.
Conductivity [µS/cm] / [mS/cm]
Resistivity [Ω·cm] / [kΩ·cm] / [MΩ·cm]
Salinity SaL [ ]
Total dissolved solids TDS [mg/l] / [g/l]
01.07.2011
08:00
Menu USB output