Cond 7310 Transmitting data (USB interface)
ba75933e04 06/2013 41
8 Transmitting data (USB interface)
8.1 Options for data transmission
Via the USB interface you can transmit data to a PC. The following table shows
which data are transmitted to the interface in which way:
Data Control Operation / description
Current mea-
sured values of
all connected
sensors
Manual With <F2>/[USB output].
Simultaneously with every manual
storage process (see section 7.1
M
ANUAL DATA STORAGE, page 35).
Automatic, at
intervals
With <F2__>/[USB output].
Then you can set the transmission
interval.
Simultaneously with every auto-
matic storage process (see section
7.2 A
UTOMATIC DATA STORAGE AT
INTERVALS, page 36).
Stored mea-
sured values
Manual Displayed dataset with <F2>[USB
output] after calling up from the
storage.
All datasets with the Output to USB
function.
Details (see section 7.3.1 D
ISPLAYING
AND EDITING THE MEASUREMENT DATA
STORAGE, page 38).
Calibration
records
Manual Displayed calibration record with
<F2>/[USB output].
Details (see section 5.2.4 C
ALIBRA-
TION DATA, page 25).
Automatic At the end of a calibration proce-
dure.
The following rule applies: With the exception of the menus, shortly
pressing the [USB output]<F2> key generally outputs the display
contents to the interface (displayed measured values, measure-
ment datasets, calibration records).