2.4 COMBINE A GROUP OF INSTRUMENTS USING THE FUNCTION "JOIN"
The function "JOIN" enables the grouping of instruments through
wireless data transmission. The grouping prevents the accidental
reading of measuring values from instruments of another measuring
group. The function "JOIN" can be performed using the cable
connection or with wireless data transmission.
The function "JOIN" through cable connection joins all the
instruments connected to a group. Previously used groupings are
cancelled.
For the function "JOIN" in wireless mode only one instrument can be added to the group at a time. No instruments
must be connected by cables at the same time.
2.4.1 PROCEDURE "JOIN" WITH CABLE CONNECTION
1. Connect tall instruments (BlueLEVEL BASIC and
BlueMETER BASIC) to be joined to a group with
the cables provided and switch on all the
instruments
2. Select on the BlueMETER BASIC the menu
"JOIN" using the <ON/MODE> key. Confirm with
the <ENTER> key. All instruments connected are
now searched and joined to a group.
3. After establishing the group a "REFRESH" will be
performed.
4. After the grouping the command "SENSOR" will
be performed. As the measuring mode is still to be
defined, the message "not-defined" will be
displayed
5. The measuring mode ("A", "B", "A B" or "A-B" must now be selected. Select the mode preferred using the
<ZERO/SELECT> keys (corresponding to the menu [SENSOR]) and confirm with <ENTER>
It is also possible to execute any other function of the menu. For instance it is possible to integrate with
further "JOIN" commands additional instruments with wireless connection into the group
6. The sensors must be selected. Select the sensors for A and - depending on the measuring mode - also for
B using the <ZERO/SELECT> keys and confirm with <ENTER>. The measuring values are now shown in
the display according to the selected configuration.
7. After a successful grouping on both instruments the green LED "READY" will blink shortly as many times
as instruments are joined in the measuring group (including the own address)
8. For using the wireless mode (the wireless mode must be switched-on on each instrument) the cables can
now be removed. After removing the cables the measuring values will be "freezed" for a short while and
replaced by empty zeroes until, after successful connection, the measuring values will be displayed again
After successful connection the blue LED "LINK" will be lighting on all the instruments connected.
The procedure "JOIN" through cables cancels and overwrites all previously registered groups in the BlueMETER
BASIC as well as in the BlueLEVEL BASIC.