page 40 of 84
A4 BASIC SETUP OF THE INCLINATION MEASURING INSTRUEMNTS / PLUS - MINUS RULE
The instrument should always be placed in a way that the cable output is on the right hand side.
Inclined to the right: positive
Declined to the right: negative
(Practical remark: positive values
indicate that some material may be
removed on the measured piece)
Before starting the measurement enough acclimatisation time must be allowed for the instruments.
Place the instrument carefully to the required measuring spot. (Surface) and read the value on the BlueMETER
BASIC.
It is important to measure always in the direction of the cable connector, respectively the cable.
Example: Flatness measurement of a surface plate
IMPORTANT
:
• Touch the instruments only at the
handle (Temperature!!!)
• Measuring from left to right and from
close to far.
• Cable in measuring
direction