page 46 of 84
After the first reading the following display is seen:
This display means the first reading was successful. The instrument must be turned 180 degrees without lifting and
slid exactly onto the previously marked position
Use the zapper or confirm with <ENTER>. On the display the progress of the collected values are again visible in
graphic form.
When the curve is getting a straight line confirm again the second value with the zapper or with <ENTER>
After the second reading the following display is seen:
With this the reversal measurement is finished and the instrument shows the true absolute value.