SpectraLight QC
16
2. Use the up (▲) and down (▼) arrow switches to highlight the desired option.
3. After the option is highlighted, press the EDIT switch to access the editing screen.
4. Enter required text or select desired option for the owner options. Refer below for additional information on
editing the owner options.
Entering text using the front panel:
Use the left (), right (►), up (), and down () arrow switches to highlight the letter/number in the list.
Use the BACK switch to delete one character at a time.
Press the ENTER switch to add the character to the name.
Continue with additional letter/numbers until the name is complete.
Press the DONE switch and then the Yes switch when finished.
Entering text using the application:
Make sure the editing window is displayed and enter required name (15 characters maximum).
Press Enter key on the keyboard when finished.
5. When finished with owner options, press the EXIT switch to exit the Owner screen.
Primary System Settings (SYS)
1. Press the SYS switch to access the System Settings screen.
The following are descriptions of the system settings.
System
Settings
Value or option
selection
Factory
Default
Description
DL + UV
Auto, Manual
Auto
Automatic add UV to daylight or manually add UV
to daylight.
Dim Reference
Voltage, Lux
Lux
Set fluorescent lamp dim control by lux or by
voltage. UV is always dimmed by voltage.
Filter
D50, D65
D50 or D65
(depending
on model)
NOTE: The filter is set at the factory for a specific
model. This setting should only change if the
installed filter is changed by an authorized service
technician.
Up/Down keys
Dim Control, Tri Lux,
LCD Control
Dim Control
Set up and down switches to be used for Dim
control, Tri Lux or LCD brightness/contrast
control.
Acoustic
Signal
Enabled, Disabled
Disabled
Enable or disable an audible signal to occur when
a message appears on the front panel.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the X-Rite Macbeth Lighting SpectraLight QC and is the answer not in the manual?

X-Rite Macbeth Lighting SpectraLight QC Specifications

General IconGeneral
BrandX-Rite
ModelMacbeth Lighting SpectraLight QC
CategoryLighting Equipment
LanguageEnglish

Related product manuals