<4. Wiring>
34
IM 01E24A01-01EN
(4) Waterproof Gland with JIS G3/4 Female
Remote Sensor
Optional code: EW
Gasket
Washer
* When connecting G1/2, remove the conversion plug and
connect directly to cable entry.
Cable
G3/4
G1/2
Conversion plug (x2)*
F0417.ai
When working on conduit or flexible tube (G3/4)
Remote Transmitter (AXG1A/AXFA11)
Optional code: EW5 (AXG1A), EW (AXFA11)
Gasket
Washer
* When connecting G1/2, remove the conversion plug
and connect directly to cable entry.
When working on conduit or
flexible tube (G3/4)
Extension plug (x2)
Cable
G3/4
G1/2
Conversion plug (x5)*
F0418.ai
4.3.2 Conduit Wiring
When wiring the conduits, utilize the waterproof gland to
prevent water from owing in through the conduit. Place
the conduit pipe on an angle as shown in the following
gures.
Install a drain valve at the low end of the vertical pipe, and
open the valve regularly.
To prevent noise, please do not use the same conduit for
signal cable and power cable.
Drain valve
F0419.ai
Figure 4.3.1 Integral Flowmeter
Drain valve
F0420.ai
Figure 4.3.2 Remote Sensor
F0421.ai
Drain valve
Figure 4.3.3
Remote Transmitter (AXW4A)
F0422.ai
Drain valve
Figure 4.3.4
Remote Transmitter (AXG1A, AXFA11)

Other manuals for YOKOGAWA AXW4A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA AXW4A and is the answer not in the manual?

YOKOGAWA AXW4A Specifications

General IconGeneral
BrandYOKOGAWA
ModelAXW4A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals