7-14
IM CA500-01EN
SF Sets or queries the source function
Command = SFm<CRLF> -> Return = SFm<CRLF>
Command = SF?<CRLF> -> Return = SFm<CRLF>
Parameters
m = 0: DCV
1: DCA
2: Ω
3: TC
4: RTD
5: Freq
7: OFF
Use the PU command to set the amplitude and pulse count of the PULSE source.
SH Sets or queries the value corresponding to 100% of FUNCTION2
Command = SHm<CRLF> -> Answer = SHm<CRLF>
Command = SH?<CRLF> -> Return = SHm<CRLF>
Parameters
m = 100% value
The setting range and resolution are the same as the display range of the selected source range.
SL Sets or queries the value corresponding to 0% of FUNCTION2
Command = SLm<CRLF> -> Answer = SLm<CRLF>
Command = SL?<CRLF> -> Return = SLm<CRLF>
Parameters
m = 0% value
The setting range and resolution are the same as the display range of the selected source range.
SO Starts or stops or queries the source
Command = SOm<CRLF> -> Answer = SOm<CRLF>
Command = SO?<CRLF> -> Return = SOm<CRLF>
Parameters
m = 0: Source stop
1: Source start
SP Sets or queries the pulse count operation state
Command = SPm<CRLF> -> Answer = SPm<CRLF>
Command = SP?<CRLF> -> Return = SPm<CRLF>
Parameters
m = 0: Stop
1: Start