7-15
IM CA500-01EN
1
2
3
4
5
6
7
Index
USB Function
SR Sets or queries the source range
Command = SRm<CRLF> -> Answer = SRm<CRLF>
Command = SR?<CRLF> -> Return = SRm<CRLF>
Parameters
[DCV] m = 0: 100 mV
1: 5 V (1 V on the CA150)
2: 30 V (10 V on the CA150)
3: 30 V
4: 1-5 V
5: 5 V
6: 1-5 V √
[DCA] m = 0: 20 mA
1: 4-20 mA
2: 4-20 mA Simulate
3: 4-20 mA
[Ω] m = 0: 400 Ω (500 Ω on the CA150)
1: 4000 Ω (5 kΩ on the CA150)
[TC] m = 0: K 1: E 2: J
3: T 4: R 5: B
6: S 7: N 8: L
9: U 10: C 11: XK
12: A 13: D 14: G
15: Platinel II 16: PR20-40
[RTD] m = 0: PT100 1: JPT100 2: PT100 (3850)
3: PT100 (3926) 4: R 5: B
6: S 7: N 8: L
9: U 10: C 11: XK
12: A 13: D 14: G
[PULSE] m = 0: 500 Hz (100 Hz on the CA150)
1: 5000 Hz (1000 Hz on the CA150)
2: 50 kHz (10 kHz on the CA150)
3: 50 kHz
4: CPM
SSC Sets or queries the FUNCTION2 display
Command = SSCm<CRLF> -> Answer = SSCm<CRLF>
Command = SSC?<CRLF> -> Return = SSCm<CRLF>
Parameters
m = 0: Normal management display
1: Percentage display
7.4 Commands