7-16
IM CA500-01EN
TC Sets or queries the pulse count time
Command = TCm<CRLF>  -> Answer = TCm<CRLF>
Command = TC?<CRLF>  -> Return = TCm<CRLF>
Parameters
  m = 1 to 60
TE Sets or queries the TC or RTD (source) display
Command = TEm<CRLF>  -> Answer = TEMm<CRLF>
Command = TE?<CRLF>  -> Return = TEm<CRLF>
Parameters
  m =  0: Temperature value
    1: mV value (resistance)
    2: Room temperature
ERR13 results when the function is not set to TC or RTD.
TS Executes manual saving
Command = TS<CRLF>  -> Answer = TS,OK<CRLF>
TT Sets or queries the international temperature standard
Command = TTm<CRLF>  -> Answer = TTm<CRLF>
Command = TT?<CRLF>  -> Return = TTm<CRLF>
Parameters
  m =  0: IPTS-68
    1: ITS-90 (default value)
UP Increases the mth digit of the source value by 1.
Command = UPm<CRLF>  -> Answer = UP, OK<CRLF>
Parameters
  m =  1 to 5 (1: least significant digit to 5: most significant digit)
7.4  Commands