-49-
Problem Possible Cause Solution
d Shutter LED
stay lighted
against manual
switch-off or
command off
Shutter external
control signal is at
High
Set shutter external signal to Low, or stop signal input
e Shutter LED
blinks(Repeat
blinking twice
and off cycle)
Interlock is working Connect interlock key if not
Attach DM block cover properly
f Shutter LED
blinks(Repeat
blinking 5
times and off
cycle)
Total on/off cycle of
shutter surpassed
beyond guaranteed
60,000 times
Blinking stops by manual button.Shutter works
accordingly
Please call you local rep and exchange the shutter if
this problem occurs