CAUTION
Please do not install the unit under the following conditions.
• Under direct sunlight
• Close to heating objects
• Polluted with oily smoke, steam, dusts or corrosive gas
• Close to any objects with heavy electromagnetic field
• Close to high-voltage apparatus or a power line
• Under heavy mechanical vibration
• Unstable place
<Installation>
Please install the unit horizontally.
.
3.3.2 Connection
3.3.2.1 CSUX1CU-F1 Model
(1) Connect EXT CONTROL / INTERLOCK of CSU-X1 with the CSU connector of
control unit ;CSUX1CU-F1, with supplied cable.
(2) Connect controller interlock key to EXT INTERFACE1/I.LOCK.
(3) For communication, connect RS232C1 connector of control unit with PC using
supplied RS232C cable.
(4) To use Filter Wheel, connect FW1 and MOTOR1 connector with supplied FW1
cable.
(5) To connect second FW, connect FW2 and MOTOR 2 connector with supplied FW2
cable.
(6) After checking the power switch is OFF, connect the power cable
Please use the power receptacle which fulfills the following conditions.
* 3 pole electric socket equipped with protective earth terminal.
* Rated supply voltage:100~240VAC
* Power voltage frequency range:90~264VAC
* Rated supply frequency:50~60Hz
* Power frequency latitude:48~63Hz
* Maximum electric power consumption:200VA