1-14
IM 701610-61E
Address Trigger
Trigger Only on the Address Pattern
Trigger Condition
Address Pattern: A4
Data Pattern: Not applicable
Condition: True
Ignore the Restart Condition: No
Ignore Unexpected Start/Stop Condition: No
S
25 AE 57
27
FE
98 99 27 P
A4
Address + R/W bit
Match against the specified address pattern.
Trigger activated here.
Trigger Only on the Data Pattern
Trigger Condition
Address Pattern: Not applicable
Data Pattern: 27
Condition: True
Ignore the Restart Condition: No
Ignore Unexpected Start/Stop Condition: No
Position = Anywhere
Address + R/W bit
S
25 AE 57
27
FE
98 99 27 P
A4
2. Trigger activated here
1. Compare until a match occurs against the specified pattern (27)
Position = First Byte
S
27 AE 57
27
FE
98 99 27 P
A4
2. Trigger activated here
1. Compare against the specified pattern (27). A match
Address + R/W bit
1.4 Trigger Setting Examples