EasyManua.ls Logo

YOKOGAWA DL1640

YOKOGAWA DL1640
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1-2
IM 701610-61E
1.2 Connecting the Probe
Input Terminals
Connect the probe to one of the input terminals located at the lower section of the front
panel. The input impedance is 1 M±1.0% and approximately 28 pF.
CH 1 CH 2 CH 3 CH 4
1M 28pF 300V
CAT
WARNING
To prevent fire or electric shock, do not use this instrument for category
II, III, or IV measurements.
CAUTION
The maximum input voltage for 1 M input is 300 VDC or 300 Vrms
when the frequency is 1 kHz or less. Applying a voltage exceeding this
maximum can damage the input section. If the frequency is above 1
kHz, the input section may be damaged even when the voltage is below
this value.
Precautions to Be Taken When Connecting a Probe
When using the I
2
C-bus trigger function, apply the SCL (serial clock) signal and SDA
(serial data) signal to the CH1 and CH2 input terminals, respectively.
When connecting a probe to the instrument for the first time, perform phase correction
of the probe as described in section 3.5, “Compensating the Probe (Phase
Correction)” in the
DL1620/DL1640/DL1640L User’s Manual IM 701610-01E
. Failure
to do so may result in unstable gain across different frequencies, thereby preventing
correct measurement. Calibration must be performed for each channel.
Note that if the object being measured is directly connected to the instrument without
using a probe, correct measurements may not be possible due to loading effects.
Note
The data analysis function and the data search function of the I
2
C bus can be used against
the CH3 and CH4 signals. For details, see section 1.5, “Analyzing/Searching Data.”

Related product manuals